Dosimetric validation of Acuros® XB algorithm for RapidArc™ treatment technique: A post software upgrade analysis


1 Department of Applied Science and Humanities, Dr. A.P.J Abdul Kalam Technical University, Lucknow, Uttar Pradesh; Department of Radiation Oncology, Division of Medical Physics, Rajiv Gandhi Cancer Institute and Research Centre, New Delhi, India
2 Department of Radiation Oncology, Division of Medical Physics, Rajiv Gandhi Cancer Institute and Research Centre, New Delhi, India
3 Department of Applied Science and Humanities, Bundelkhand Institute of Engineering and Technology, Jhansi, Uttar Pradesh, India
4 Department of Radiotherapy, Mahatma Gandhi Memorial Medical College, Indore, Madhya Pradesh, India

Correspondence Address:
Lalit Kumar,
Department of Radiation Oncology, Division of Medical Physics, Rajiv Gandhi Cancer Institute and Research Centre, Sector-5, Rohini, New Delhi - 110 085
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None

DOI: 10.4103/jcrt.JCRT_1154_19

Aim: To validate the Acuros® XB (AXB) algorithm in Eclipse treatment planning system (TPS) for RapidArc™ (RA) technique following the software upgrades. Materials and Methods: A Clinac-iX (2300CD) linear accelerator and Eclipse TPS (Varian Medical System, Inc., Palo Alto, USA) was used for commissioning of AXB algorithm using a 6 megavolts photon beam. Percentage depth dose (PDD) and profiles for field size 2 cm × 2 cm, 4 cm × 4 cm, 6 cm × 6 cm, 10 cm × 10 cm, 20 cm × 20 cm, 30 cm × 30 cm to 40 cm × 40 cm were taken. AXB calculated PDDs and profiles were evaluated against the measured and analytical anisotropic algorithm (AAA)-calculated PDDs and profiles. Test sites recommended by American Association of Physicists in Medicine task group (AAPM TG)-119 recommendation were used for RA planning and delivery verification using AXB algorithm. Results: Dosimetric analysis of AXB calculated data showed that difference between calculated and measured data for PDD curves were maximum <1% beyond the depth of dose maximum and computed profiles in central region matches with maximum <1% for all considered field sizes. Ion-chamber measurements showed that the average confidence limit (CLs) was 0.034 and 0.020 in high-gradient and 0.047 and 0.042 in low-gradient regions, respectively, for AAA and AXB calculated RA plans. Portal measurements show the average CLs were 2.48 and 2.58 for AAA and AXB-calculated RA plans, with gamma passing criteria of 3%/3 mm. Conclusions: AXB shows excellent agreement with measurements and AAA calculated data. The CLs were consistent with the baseline values published by TG-119. AXB algorithm has the potential to perform photon dose calculation with comparable fast calculation speed without negotiating the accuracy. AAPM TG-119 was successfully implemented to access the proper configuration of AXB algorithm following the TPS upgrade.


Print this article
Search
 Back
 
  Search Pubmed for
 
    -  Kumar L
    -  Bhushan M
    -  Kishore V
    -  Yadav G
    -  Gurjar OP
 Citation Manager
 Article Access Statistics
 Reader Comments
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed256    
    PDF Downloaded3    
    Cited by others 1    

Recommend this journal